4
Maintenance
We recommend a standard bearing change after 15,000 operating hours at
the latest. Moreover, we are recommending an exchange of the rotor unit
after 45,000 operating hours at the latest.
Such maintenance work can only be done by the Oerlikon Leybold Vacuum
Service. If required contact the Oerlikon Leybold Vacuum service center nea-
rest to your location. You can find the address on our internet page
www.oerlikon.com.
At high pump loads - for example during cyclic operation, at high gas
throughputs or at high ambient temperatures - the aforementioned mainten-
ance work should be carried forward.
When an adsorption trap is used, regenerate or renew the adsorption agent
regularly; refer to the operating instructions provided with the trap.
The frequency converter is maintenance free.
4.1
Cleaning
If required clean the turbomolecular pump of dust with a dry cloth.
4.2
Oerlikon Leybold Vacuum Service
Whenever you send us in equipment, indicate whether the equipment is con-
taminated or is free of substances which could pose a health hazard. If it is
contaminated, specify exactly which substances are involved. You must use
the form we have prepared for this purpose.
A copy of the form has been reproduced at the end of these Operating
Instructions: “Declaration of Contamination for Compressors, Vacuum Pumps
and Components”. Another suitable form is available from www.oerlikon.com
→
Oerlikon Leybold Vacuum
→
Documentation
→
Download Documents.
Attach the form to the equipment or enclose it with the equipment.
This statement detailing the type of contamination is required to satisfy legal
requirements and for the protection of our employees.
We must return to the sender any equipment which is not accompanied by a
contamination statement.
Maintenance
54
130000761_002_A0 - 08/2008 - © Oerlikon Leybold Vacuum
Contamination
Form