Applications which can be implemented with the aid of the serial interface:
Application
Benefits to the customer
How to do it
Networking of several pumps
Savings relating to the costs for
With Field Bus systems like Profibus
and other equipment
signalling cables
Automation
Savings related to repetitive manual work For example by a control computer
Avoidance of warnings and
■
Precise planning for
Monitoring of:
warnings before overload
maintenance
■
Motor current P5
operation and early detection
■
Improved reliability of sensitive
■
Motor temperature P7
of a failing pump
production processes in a vacuum
■
Frequency converter temperature P11
Standby operation
■
Extending the service life
Reducing the rotor’s frequency
for the ball bearings
through P24
■
Cutting energy consumption
Troubleshooting
Quick analysis of problems
Reading of error memories
P171, P174 and P176:
error code, speed, operating hours for error
Slow pressure control by
Dispensing with a flow controller
Changing the rotor frequency
changing the pumping speed
through parameter 24
Reducing the maximum motor
Cost savings through smaller
With P139, motor current
current
power supply units if peak loads
reduction factor
can be reduced
Starting the pump with a delay if
Cost savings through smaller
With P36, delay
several consumers are
power supply units
connected to the same PSU
if peak loads can be reduced
Frequency converter as a simple
Dispensing with pressure gauges
Monitor motor current P5; second function
pressure gauge, since
for “Normal Operation” relay: relay switches
motor current is dependent
as soon as the motor current threshold
on the vacuum conditions
is tripped.Adjust second function: P29
Set motor current thresh.: P27
Lowering the normal operation
Normal operating mode is attained
Reduce frequency threshold
threshold
faster, processes can be started faster
through P25
Interfaces
40
130000761_002_A0 - 08/2008 - © Oerlikon Leybold Vacuum
3.3
Interfaces
The frequency converter has a RS 232 interface as standard (SERVICE X5)
and is optionally equipped with serial interfaces:
■
RS 485 C
■
Profibus DP
■
Ethernet/IP
The TURBO.DRIVE 400 is configured through the parameters according to
the parameter list. Pxxx denotes parameter value xxx.
The PC software "TURBO.DRIVE Server" allows convenient access by the
user to the parameters of the frequency converter.
In the case of BETA units simultaneous operation of the RS 232 interface in
parallel to the RS 485 interface or the Profibus interface is not yet reliably
supported. For this reason only one interface at a time should be connected
to the unit.