Installation
37
130000761_002_A0 - 08/2008 - © Oerlikon Leybold Vacuum
24 V
5 V
1 k
Ω
1 k
Ω
TURBO.DRIVE 300
7
8
6
24 V
TURBO.DRIVE 300
7
8
6
0 V = STOP
24 V = START
24 V
TURBO.DRIVE 300
7
8
6
Fig. 2.19 Pin assignment of the REMOTE (X1) connector
Relay functions
8 7 6
5 4 3 2 1
LEYBOLD AG
WERK KÖLN
LEYBOLD
GA 06.215
21.11.94
Fa. ESCH
00.00.S.001
9er Buchsenbelegung
9
Pin assignment of the connector
24 V
5 V
2,6 k
Ω
3,6 k
Ω
TURBO.DRIVE 400
7
8
6
24 V
TURBO.DRIVE 400
7
8
6
0 V = STOP
24 V = START
24 V
TURBO.DRIVE 400
7
8
6
7,2 k
Ω
6,2 V
Pin assignment for the Start/Stop input
Switching threshold for
the Start/Stop control
input:
Low-Level: < 9 V
High-Level: > 10,5 V
Start/Stop operation
Example 1: Operation via a PLC
Contact open = STOP
Contact closed = START
Example 2: Operation via contacts
TURBO.DRIVE S
5
3
4
n. c.
n. o.
com.
2
9
1
n. c.
n. o.
com.
Relay - Normal operation
■
While deceleration, acceleration, Stop:
4 connected to 5 (as shown; passive)
■
During normal operation (f > 0,9·f
nom.
):
4 connected to 3 (active)
Relay - Error
■
No error: 1 connected to 2 (as shown; passive)
■
Error is present: 1 connected to 9 (active)
TURBO.DRIVE 400
Relay output rating: 42 V, 0.5 A
Caution: No power feed
is allowed at pin 7
TURBO.DRIVE 400
TURBO.DRIVE 400