Note: The device must be switched off before removing the battery.
8.
To remove the battery, put the device on a soft, flat surface. Move the battery lock to the
unlocked
position, and then slowly pull the battery out of the unit.
9.
To install the probe, find the red dots on the main unit port and on the plug at the end of the
probe cord. See Figure 6-3. Align the two red dots and then plug the probe into the main unit.
See Figure 6-4.
Note: Do not scratch the surface of the probe. This will result in inaccurate readings. Only the probe
supplied by the original manufacturer may be used with this device. Please contact Life Systems, Inc.
in the event a replacement probe is needed.
Figure 6-2
Back view of main unit
with battery in place
Figure 6-3
Probe Installation
Figure 6-4
Probe Installed
Page
17