After switching on the device, enter the system reset screen by selecting Phantom in the
Scan Mode before returning to Ready screen.
A3.
Enter pre-scan state by pressing SCAN key (FigureA-3).
A4.
Adjust the distance between probe and phantom; let the round dark area (target) be in the
middle of fan-image in pre-scan. The red line shown in Figure A-3 is the central line of the measurement.
The object being measured must be across this center line
A5.
Enter the Scan screen (see Figure A-4) by pressing the SCAN key
A6.
Take 5 separate measurements of the phantom to confirm the accuracy of the scanner.
Calibration is not necessary if results are within 128±10ml. (or the standard volume for the phantom
being used)