7.4.8 Volume Calibration
A “phantom bladder” will be needed for calibration. The phantom mimics a human bladder with a
known volume. The phantom should be scanned and the measured volume should be the same as the
phantoms standard volume. If the measured volume is different from phantom’s standard volume,
use the correcting parameter to calibrate the device. The range of its correction is ±40ml.
1.
Press the
▲
/
▼
keys or touch Volume Calibration. Press
or touch Volume Calibration to
enter password (see Figure 7-8)
2.
Once you have entered the password, press
or touch
to move to the calibration screen
(see Figure 7-9)
3.
Press the
▲
/
▼
keys again to set ‘+’ ‘-‘ , then press
▲
/
▼
to set the calibration value
4.
Press
key or touch
to save the calibration value. Press the
key or touch
to
exit the calibration screen
5.
Scan Images: select the number of images to scan
a.
Press the
▲
/
▼
keys or touch
Scan Images
b.
Press the
key to underline the numbers of images you need to scan.
c.
Press the
▲
/
▼
keys or touch the number of images (12 or 24) to scan
d.
Press the
key or touch
Scan Images
to save the setting
Figure 7-8
Password Input Screen for Calibration
Figure 7-9
Calibration Screen