Use only with IRMA monitor supporting MASIMO airway adapter.
After IRMA adult / child model airway adapter access to circuit will increase 6mL of
dead space, and therefore prohibited for infants.
Do not use IRMA infant airway adapter for adults, otherwise it will lead to excessive
flow resistance.
To avoid the patient’s secretions clogging the adapter window caused operational
errors, do not put IRMA airway adapter between the endotracheal tube and elbows.
Ensure IRMA module is placed in the vertical direction and the status LED upward.
To avoid effect the light propagation of airway adapter windows, do not use airway
adapter andquantitative spray or spray together.
Ensure IRMA module used in electromagnetic environment that this specification
defined, in case of to be affected by mobile or portable communication devices.
IRMA module is only designed as an auxiliary method to patient assessment. Always
with other vital signs and symptoms assessed with the use of equipment.Please use it
together with other vital signs and symptoms assessed equipment.
Do not stretch IRMA module probe cable.
7.2.2 Side stream
ISA side stream CO2 module is through the sampling tube to collect a small quantity of
mixed gas from the endotracheal tube of breathing circuit, which is transferred to the
sampling gas input port of the monitors, the module will measure and analysis the sample
gas.Nomoline sampling tube has a special impermeable area regardless of the direction, the
impermeable area with a hydrophobic bacterial filter and made of special polymer.
Impermeable area can remove the condensate that produced by sampling gas or removal
inhaled water, serve to prevent cross-contamination caused by moisture intruded into the
monitor. The design of sampling tube ensures sampling air flow continuous unimpeded, and
therefore the response time of measure the gas is very short.
NOTE
Side stream ISA module can only be used by trained or authorized medical staff.
ISA module can not be used in flammable anesthetic gas environment.
Use only with ISA module supporting MASIMO sampling tube.
To reduce the risk of bridling or entangleding the patient, be sure to carefully straighten
out the sampling tube.
Summary of Contents for Crius V6
Page 1: ...1 182 Crius V6 Ventilator Operator s Manual...
Page 15: ...15 182 FOR YOUR NOTES...
Page 26: ...26 182 FOR YOUR NOTES...
Page 28: ...28 182 2 2 Equipment Appearance 2 2 1 Front View 1 13 9 10 12 11 8 7 6 5 2 4 3...
Page 32: ...32 182 FOR YOUR NOTES...
Page 44: ...44 182 FOR YOUR NOTES...
Page 56: ...56 182 FOR YOUR NOTES...
Page 88: ...88 182 FOR YOUR NOTES...
Page 100: ...100 182 FOR YOUR NOTES...
Page 125: ...125 182 FOR YOUR NOTES...
Page 149: ...149 182 FOR YOUR NOTES...
Page 156: ...156 182 FOR YOUR NOTES...