174
/
182
Trend
log
Factory default
setting
Graphic
trend-interval
1s
Tabular
trend-interval
1
s
Alarm log-level
All
15.11.6 Therapy
O
2
Therapy
Factory default
setting
Adu
Ped
Neo
O
2
%
21 %
21 %
21 %
Flow
5.0
L/min
3.0
L/min
2.0L/min
Pmax
40cmH
2
O
30cmH
2
O
20cmH
2
O
15.12 EMC and Radio Regulatory Compliance
15.12.1 EMC
Crius V6 ventilator is in compliance with IEC 60601-1-2 for EMC.
The essential performance verified during the immunity testing comprise of TVi control
accuracy, TVi monitoring accuracy, CO
2
monitoring accuracy, O
2
control accuracy, O
2
monitoring accuracy, PEEP control accuracy and PEEP monitoring accuracy.
NOTE
Using accessories other than those specified may result in increased electromagnetic
emission or decreased electromagnetic immunity of the equipment.
The ventilator or its components should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the ventilator or its
components should be observed to verify normal operation in the configuration in
Summary of Contents for Crius V6
Page 1: ...1 182 Crius V6 Ventilator Operator s Manual...
Page 15: ...15 182 FOR YOUR NOTES...
Page 26: ...26 182 FOR YOUR NOTES...
Page 28: ...28 182 2 2 Equipment Appearance 2 2 1 Front View 1 13 9 10 12 11 8 7 6 5 2 4 3...
Page 32: ...32 182 FOR YOUR NOTES...
Page 44: ...44 182 FOR YOUR NOTES...
Page 56: ...56 182 FOR YOUR NOTES...
Page 88: ...88 182 FOR YOUR NOTES...
Page 100: ...100 182 FOR YOUR NOTES...
Page 125: ...125 182 FOR YOUR NOTES...
Page 149: ...149 182 FOR YOUR NOTES...
Page 156: ...156 182 FOR YOUR NOTES...