In PC-SIMV mode, you need to set the following basic ventilation parameters:
1.
[Pmax]: Can prevent high airway pressure caused
by
patient
2.
[PEEP]: Positive end-expiratory pressure
3.
[Psupp]: Pressure support leve
4.
[Pinsp]: Inspiration pressure leve
5.
[Tinsp]: Inspiration time
6.
[Rate]:
7.
[Tslope] Breathing frequency
8.
[Trigger]: Inspiration trigger level
9.
[Exp%]: Expiration trigger level
10.
[
O2
%]: Oxygen concentratio
n
In PC-ACV mode, you can set the following sigh function parameters as required:
1.
[
Sigh
]: Switch for turning on sigh function
2.
[
Interval
]: Number of sigh cycles
3.
[
Delta Pinsp
]:
Inspiration pressure of sigh cycles
6.5.8 PVC-SIMV
PVC-SIMV is volume-synchronized intermittent mandatory ventilation mode. It provides the
minimum number of mandatory breaths based on the preset intermittent mandatory ventilation
frequency. The provided mechanical ventilation mode is volume mode (PRVC mode). If patient
triggers within the trigger window, ventilator delivers mandatory PRVC breath once. Mandatory
Summary of Contents for Crius V6
Page 1: ...1 182 Crius V6 Ventilator Operator s Manual...
Page 15: ...15 182 FOR YOUR NOTES...
Page 26: ...26 182 FOR YOUR NOTES...
Page 28: ...28 182 2 2 Equipment Appearance 2 2 1 Front View 1 13 9 10 12 11 8 7 6 5 2 4 3...
Page 32: ...32 182 FOR YOUR NOTES...
Page 44: ...44 182 FOR YOUR NOTES...
Page 56: ...56 182 FOR YOUR NOTES...
Page 88: ...88 182 FOR YOUR NOTES...
Page 100: ...100 182 FOR YOUR NOTES...
Page 125: ...125 182 FOR YOUR NOTES...
Page 149: ...149 182 FOR YOUR NOTES...
Page 156: ...156 182 FOR YOUR NOTES...