182
Pmean
-5 to +105
mbar
± (2 + 4% of actual)
Pplat
-5 to +105
mbar
± (2 + 4% of actual)
PEEP
0 to 40
mbar
± (2 + 4% of actual)
Ptracheal
-10 to 110
mbar
± (5 + 10% of actual)
Poutput
-10 to 110
mbar
± (5 + 10% of actual)
Pambient
0 to 3000
mbar
± (5 + 10% of actual)
P0.1
-105 to 5
mbar
± (1 or 25% of reading)
VTe mand
0 to 2500
mL
VTe mand
≦
100:± (2 + 15% of actual)
VTe mand>100:± (4 + 15% of actual)
VTe spon
0 to 2500
mL
VTi mand
≦
100:± (2 + 15% of actual)
VTe mand>100:± (4 + 15% of actual)
VTi mand
0 to 2500
mL
VTi mand
≦
100:± (2 + 15% of actual)
VTe mand>100:± (4 + 15% of actual)
VTi spon
0 to 2500
mL
VTi spon
≦
100:± (2 + 15% of actual)
VTi spon>100:± (4 + 15% of actual)
MVe
0 to 100
L/min
± (0.2 + 10% of actual)
MVe spon
0 to 100
L/min
± (0.2 + 10% of actual)
MVi mand
0 to 100
L/min
± (0.2 + 10% of actual)
MVi spon
0 to 100
L/min
± (0.2 + 10% of actual)
Tinsp
0 to 100
s
± (0.05 + 5% of actual)
Texp
0 to 100
s
± (0.05 + 5% of actual)
I:E
/
±50 ms or ±6%, whichever is greater
PIF
0 to 200
L/min
± (2 + 15% of actual)
PEF
-200 to 0
L/min
± (2 + 15% of actual)
Rtotal
0 to 250
bpm
± (5 or 10% of actual)
RR spon
0 to 250
bpm
± (5 or 10% of actual)
FiO2 Cal
21 to 100
Vol.%
± (2.5 + 2.5% of actual)
FiO2 Measu
21 to 90
Vol.%
When TV=500ml, f=10 bpm, I:E=1:2,
≤
50s;
When TV=150ml, f=20 bpm, I:E=1:2,
≤
80s;
When TV=30ml, f=30 bpm, I:E=1:2,
≤
120s
Rlung
0 to 300
mbar/
(
L/s
)
1.0
mbar/
(
L/s
)
~20 mbar/
(
L/s
)
: ±10 mbar/
(
L/s
)
2.20 mbar/
(
L/s
)
~300 mbar/
(
L/s
)
: ±(50%
of actual)
Cdynamic
0 to 100
mL/mba
r
± (5 + 20% of actual)
RSBI
0 to 10000
bpm/mi
n/L
± (10 or 25% of reading)
PTP
0 to 10
mbar*s
± (0.1 or 25% of reading)
Alarm Settings
Parameter
Setting
range
Step length
Remarks
Paw
High
limit
0 to 98
1mbar
The upper limit is
Summary of Contents for Crius V6
Page 1: ...1 182 Crius V6 Ventilator Operator s Manual...
Page 15: ...15 182 FOR YOUR NOTES...
Page 26: ...26 182 FOR YOUR NOTES...
Page 28: ...28 182 2 2 Equipment Appearance 2 2 1 Front View 1 13 9 10 12 11 8 7 6 5 2 4 3...
Page 32: ...32 182 FOR YOUR NOTES...
Page 44: ...44 182 FOR YOUR NOTES...
Page 56: ...56 182 FOR YOUR NOTES...
Page 88: ...88 182 FOR YOUR NOTES...
Page 100: ...100 182 FOR YOUR NOTES...
Page 125: ...125 182 FOR YOUR NOTES...
Page 149: ...149 182 FOR YOUR NOTES...
Page 156: ...156 182 FOR YOUR NOTES...