O
2
Supply
Failure
H
1. Check connection with O2 supply.
2. Check O2 supply pressure.
CO2 communication
stopped
H
CO2 module communication stopped
1. Check if the module is faulty.
2. While removing the ventilation module, unplug the
CO2 module.
3. Please contact your designated service representative.
EtCO
2
Overrange
L
Parameter measured values exceed the measurement
range (error range is included).
Perform CO2 module zeroing.
Contact your service personnel.
Oxygen sensor is not
connected
L
Oxygen sensor is not connected.
Please connect the oxygen sensor.
Please Replace
O
2
Sensor.
M
The O2 sensor is used up.
Replace the O2 sensor.
Please calibrate
O
2
sensor.
L
Calibrate the O2 sensor.
Calibrate O2 concentration.
SPO2 Comm. Stop
H
SpO2 module failure, or communication failure
between the module and the host; please restart the
device.
SPO2 Comm. Error
H
SPO2 No Sensor
L
SpO2 sensor falls off from the patient or monitor,
malfunctions, or sensor other than specified in this
Manual is used. Check the sensor mounting position,
whether the sensor is damaged or sensor type.
Reconnect the sensor or use new sensor.
SPO2 Sensor Off
L
Summary of Contents for Crius V6
Page 1: ...1 182 Crius V6 Ventilator Operator s Manual...
Page 15: ...15 182 FOR YOUR NOTES...
Page 26: ...26 182 FOR YOUR NOTES...
Page 28: ...28 182 2 2 Equipment Appearance 2 2 1 Front View 1 13 9 10 12 11 8 7 6 5 2 4 3...
Page 32: ...32 182 FOR YOUR NOTES...
Page 44: ...44 182 FOR YOUR NOTES...
Page 56: ...56 182 FOR YOUR NOTES...
Page 88: ...88 182 FOR YOUR NOTES...
Page 100: ...100 182 FOR YOUR NOTES...
Page 125: ...125 182 FOR YOUR NOTES...
Page 149: ...149 182 FOR YOUR NOTES...
Page 156: ...156 182 FOR YOUR NOTES...