![Northern Meditec Crius V6 Operator'S Manual Download Page 128](http://html1.mh-extra.com/html/northern-meditec/crius-v6/crius-v6_operators-manual_1708588128.webp)
128
/
182
Patient tubing (including
water
trap,
Y piece,
adapter)
Each
patient/weekly
②
B or C
Other
Mainstream CO
2
sensor
Each
patient/weekly
Refer to the cleaning
and
disinfection methods provided
by
the mainstream CO
2
vendor.
SPO2 Probe
Each
patient/weekly
Refer to the cleaning
and
disinfection methods provided
by
theSPO2
vendor.
Nebulizer
Each
patient/weekly
Refer
to
the
cleaning
and
disinfection methods provided
by
the nebulizer
vendor.
Humidifier
Each
patient/weekly
Refer
to
the
cleaning
and
disinfection methods provided
by
the humidifier
vendor.
Cleaning methods
(
Wipe and Bath Immersion
)
:
①
Wipe: Wipe with a damp cloth immersed in alkalescent detergent (soap water, etc.) or alcohol
solution and then wipe off the remaining detergent with a dry lint free cloth.
②
Immersion: flush with water first and then immerse it in alkalescent detergent (soap water, etc.)
(water temperature 40º
C recommended) for approximately three minutes. Finally clean with water
and dry completely.
Disinfection
methods
:
A: Wipe: wipe with a damp cloth immersed in medium- or high-efficiency detergent and then
wipe off
the remaining detergent with a dry lint free
cloth.
B:
Immersion:
immerse
it
in
medium-
or
high-efficiency
detergent
(alcohol
or
isopropyl
alcohol,
etc.) for more than 30 minutes (recommended time). Then clean with water and dry
completely.
C: Steam autoclave at 134º
C for 10 to 20 minutes (recommended time).
D: Ultraviolet radiation for 30 to 60 minutes (recommended time).
As necessary*: shorten the cleaning and disinfection intervals if the equipment is used
in dusty environment to ensure that the equipment surface is not covered by dust. Clean
and disinfect the inspiration safety valve assembly only when the patient’s exhaled gas
may contaminate the inspiratory limb. For disassembling and installation methods, refer
to
11.2
.
The table below lists the cleaning and disinfecting agents and autoclaving process that
may be used on the ventilator.
Name
Type
Ethanol
(75%)
Moderately efficient
disinfectant
Isopropanol
(70%)
Moderately efficient
disinfectant
Summary of Contents for Crius V6
Page 1: ...1 182 Crius V6 Ventilator Operator s Manual...
Page 15: ...15 182 FOR YOUR NOTES...
Page 26: ...26 182 FOR YOUR NOTES...
Page 28: ...28 182 2 2 Equipment Appearance 2 2 1 Front View 1 13 9 10 12 11 8 7 6 5 2 4 3...
Page 32: ...32 182 FOR YOUR NOTES...
Page 44: ...44 182 FOR YOUR NOTES...
Page 56: ...56 182 FOR YOUR NOTES...
Page 88: ...88 182 FOR YOUR NOTES...
Page 100: ...100 182 FOR YOUR NOTES...
Page 125: ...125 182 FOR YOUR NOTES...
Page 149: ...149 182 FOR YOUR NOTES...
Page 156: ...156 182 FOR YOUR NOTES...