44
Chapter 8
Adding markers to a measurement
45
Norsonic Nor118
User Guide
The keys to use
e keys used to enter the markers are (marker 1),
(marker 2), (marker 3) and (marker
4) since these keys are the lower most keys of the front
panel and thus easy to reach during a measurement. None
of these keys are used for other things during a measure-
ment either.
Other markers inserted by the instrument
As discussed in the side bar
No back-erase in profile mode
(a few pages back) the marker is added to the periods to
denote that these periods contain data acquired in pause
mode.
In addition, if you terminate an ongoing measurement
prematurely by pressing the key and later resume the
measurement by pressing / an marker will be
added to the period within which the / key was
pressed – i.e. we mark out the first period after resump-
tion.
P and S markers have priority
e marker 4 is a
toggle
marker, i.e. once activated it will
assign a marker to every period occurring until it is deac-
tivated again.
What happens if marker 4 has been activated while you
press the key or the key? Will it assign both
and or ? No, it won’t! and have priority and they
will be inserted instead of any other marker active by the
time or is pressed. So you’ll never have more
than one marker assigned to a profile period.
The keys
used to insert
markers.
Marker 1
has been
inserted