88
Chapter 18
Technical specifications
89
Norsonic Nor118
User Guide
Setting the amount of attenuation:
• In the Correction menu, press 1 (Corr.par) to gain
access to the correction parameter setup menu. Press
2 to gain access to the attenuation setting itself. Never
change this setting unless you know what you’re doing!
Use the to set the attenuation
value. Press twice to leave the menu. See Fig.
on the previous page for menu details
Self-noise compensation
e displayed values for -, - and -weighted levels may,
as an option (extension 20), be corrected automatically for
the self-noise of the microphone. e noise levels for each
of the spectral weightings must be specified. e correc-
tion is applied by unlinear level subtraction. e option
allows levels close to the self-noise to be displayed with
reduced level linearity error. When the level is equal to
the self-noise level, the correction is –3 dB.
Diffraction around the instrument casing
e instrument casing is designed to have low effects on
the sound measured at the microphone. e figure below
shows the measured effect of the instrument casing at
reference environmental conditions.
General I/O socket
A 15 pin socket for general / is placed on the right-hand
side of the instrument. Socket type: -
1
--
1.
The general
I
/
O
socket