18
Chapter 4
Simple sound measurements
19
Norsonic Nor118
User Guide
Instruments with time profile installed
If your instrument has the time profile installed, you
should set the resolution to the same value as the dura-
tion to avoid that the instrument starts logging the level
as a function of time.
Setting the time constant
If your instrument is
not
equipped with the optional
extension – parallel time constants, you may want to
specify the time constant to be used in the measurement.
e time constant is used for the , the
MAX
and the
MIN
measurements, but neither the
eq
, the
E
nor the
PEAK
makes use of it.
To set the time constant press the key until the
required time constant appears in the display. To see this,
be sure to operate the key until any of the functions
, the
MAX
or the
MIN
appears in the display first.
If your unit is equipped with multiple time constants,
you need not bother with this.
Instruments with multiple time constants
Instruments equipped with multiple time constants will
always employ all three time constants (, and ) for all
measurements. e time constants apply to the , the
MAX
and the
MIN
functions.
eq
,
PEAK
and
E
do not
make use of time constants. However, units configured for
German-speaking markets will also measure the
eqI
.
e presence of multiple time constants eliminates the
need for setup of the time constant.
Multiple time constants and statistics
Observe that the statistics buffers (optional extension) will
be based on sampling using time constant . is cannot
be changed by the user.
C or Z as spectral weighting network
e Nor118 has three spectral weighting functions in
addition to the filter bands. ese are -weighting and -
or -weighting. e -weighting is a replacement for the
previous or spectral weighting functions. A
problem when dealing with these functions has been that
none of them are properly defined in any standard.
e spectral weighting circuitry is flat within at least
16Hz to 16kHz (in the Nor118 it extends far beyond that
– viz. 20–20 000 Hz +0 –1 dB) and is well-defined in an
upcoming standard ( ).
e Nor118 can make use of two of the three spectral
weighting functions simultaneously – viz. -weighting
and - or -weighting.
To specify whether to use z- or c-weighting:
• Press > 1 () > 5 ( ) and
navigate in the menu as usual.