82
Chapter 18
Technical specifications
83
Norsonic Nor118
User Guide
Technical
specifications
U
, the specifications are for
the complete sound level meter Nor equipped with
microphone type Nor and microphone preamplifier
type Nor. Values are based on the sensitivity set to
the nominal value for the microphone: –. dB corre-
sponding to mV/Pa.
A microphone cable Nor1408 of length up to 20 m
may be used between the microphone preamplifier and
the instrument body without loss of performance. Longer
cables may be used if maximum sound pressure level or
frequency is reduced.
e definition of terms is based on - (In draft
form at the time of printing).
e options included in the basic instrument may
vary. Please check with your local supplier for the latest
information.
Type of instrument
Sound level meter -, , measuring
exponential time-weighted levels, integrating-averaging
levels and sound exposure levels. If ⁄ octave-band or 1⁄3
octave-band filters are installed, the instrument complies
with , ; and
.
Analogue inputs
Number of channels:
1
Input connector:
7 pin connector for Norsonic
microphone systems. ( .
1
..)
Preamplifier supply voltage:
± 15 volt, max 3 mA
Polarisation voltage:
0 V and 200 V, selectable.
Maximum input signal:
± 11 V peak
Input impedance:
More than 100 kohm, less than
650 pF
Measurement range:
0.3 mV to 7 V (RMS) in one range
corresponding to –10 dB to 137 dB with a microphone
sensitivity of 50 mV⁄Pa. e maximum peak value ±10 V
corresponds to 140 dB.
With the optional extension permitting extended
measurement range, peak values up to 150 dB may be
measured.