26
Chapter 5
Frequency analysis
27
Norsonic Nor118
User Guide
Setting C or Z as spectral weighting network
e Nor118 has three spectral weighting functions in
addition to the filter bands. ese are -weighting and -
or -weighting. e -weighting is a replacement for the
previous or spectral weighting functions. A
problem when dealing with these functions has been that
none of them are properly defined in any standard.
e spectral weighting circuitry is flat within at least
16 Hz to 16 kHz (in the Nor118 it extends far beyond
that) and it is well-defined in an upcoming standard (
).
e Nor118 can make use of two of the three spectral
weighting functions simultaneously – viz. -weighting and
- or -weighting.
To specify whether to use z- or c-weighting:
• Press > 1 () > 4 ( ) and
navigate in the menu as usual.
e -weighted value and the - or -weighted value will
appear as extra bar graphs to the right of the spectrum in
the display.
Activating the frequency mode
In order to enable frequency analysis, the frequency mode
must be activated.
Activating frequency mode:
• Press > 4 ( .) to gain access to the
frequency mode menu. Navigate in the menu as usual.
Set the bandwidth as required.
As long as the frequency mode has been
activated, frequency analysis will be
made during every measurement. is means that
the instrument will combine a “traditional” sound
level meter measurement and a real time frequency
analysis in octaves or third-octaves. Whether the
instrument is set to show the spectrum or the classic
sound level meter display will not affect the meas-
urement in any way. Likewise, going between the
two display modes during a measurement has no
effect on the measurement either.
Going to measure very high levels?
As an optional extension the Nor118 is able to meas-
ure very high sound pressure levels without changing
the microphone cartridge – see
High levels
in the
Technical specifications
for details.
Menu for the activation of the frequency mode
Press
SETUP
>1>4
to produce this
menu