88
Chapter 18
Technical specifications
89
Norsonic Nor118
User Guide
polarisation voltage is lowered from 200 V to about
70 V. A correction network is applied automatically to
compensate for the change in frequency response of the
microphone due to the lower polarisation voltage. e
nominal response for the “high level correction network”
is shown in the figure below.
Preamplifier attenuation
e instrument has the ability to correct for the attenua-
tion in the preamplifier. Typical values of the attenuation
is 0.2 dB. e correction can be set in the range 0.0 to
9.9 dB. e correction can be switched / to facilitate
applications of other transducers without preamplifier.
By lowering the polarisation
voltage, the level range can
be extended without chang-
ing the microphone cartridge
This feature shall be used in
combination with the Nor1225
cartridge only.
Note that the needed correction will depend on the type
of microphone, and shall only be applied when using
microphone cartridge type Nor1225.
Activating the high level range:
• Press > 5 (Lvl.range) to gain access to the level
range setting menu. Navigate in the menu as usual.
Activating the preamplifer attenuation:
• To activate the preamplifier attenuation press >
1 (Instr.) > 6 (Correct.) to gain access to the Correc-
tions menu. Navigate in the menu as usual and activate
the correction parameter Preamp by means of the
and keys. Do the same to deactivate. Preamplifier
attenuation activated is indicated by a
G
(for Gain) in
the lower line of the display