Instruction Manual
OCM F
page 62
OCM F - rev. 05 / 06.04.2018
>Name<
The relay output name may consist of four
characters max. which will be indicated in
main menu and in overview menu. Set the
name as described in chapter “23.1 Parame-
ter Menu “Measurement Place””.
Fig. 23-29 Limits submenu
>Level limit
relay<
Relay will energise if a level limit value (to be set) has been exceeded
and will de-energise if flow falls below a second limit value (to be set).
>Relay mode<,
>ON point<,
>OFF point<,
>ON delay<,
>OFF delay<,
>Name<
See >Flow limit relay<
>Vel. limit
relay<
Relay will energise if a velocity limit value (to be set) has been ex-
ceeded and will de-energise if flow falls below a second limit value (to
be set).
>Relay mode<,
>ON point<,
>OFF point<,
>ON delay<,
>OFF delay<,
>Name<
See >Flow limit relay<
>Temp. limit
relay<
Relay will energise if a temperature limit value (to be set) has been
exceeded and will de-energise if flow falls below a second limit value
(to be set).
>Relay mode<,
>ON point<,
>OFF point<,
>ON delay<,
>OFF delay<,
>Name<
See >Flow limit relay<
Fig. 23-30 Submenu impulse
Summary of Contents for OCM F
Page 105: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 105 Approvals and Certificates...
Page 106: ...Instruction Manual OCM F page 106 OCM F rev 05 06 04 2018...
Page 107: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 107...
Page 108: ...Instruction Manual OCM F page 108 OCM F rev 05 06 04 2018...
Page 109: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 109...
Page 110: ...Instruction Manual OCM F page 110 OCM F rev 05 06 04 2018...