Parameter Setting
OCM F - rev. 05 / 06.04.2018
page 71
24.3 I/O submenu “Relay”
This submenu indicates the conditions calculated by the transmitter to be sent to the relay. It
is distinguished between logical “OFF” and “ON”.
Fig. 24-4
Indication of relay output status
Observe the signal display
The actual output conditions of the relay contact on the output terminals are not shown
here. Only the signals received by the relays for output are indicated here.
This menu cannot be used to detect and to view external incorrect connections.
24.4 I/O submenu “Data storage / USB”
The >Data storage / USB< menu allows access to all measurement values saved internally.
Fig. 24-5
Submenu Data storage / USB
>Info<
Number of measurement data sets including the recording time.
>Erase<
Erase internal readings memory.
System PIN required.
>USB stick<
Transmission of measurement data to USB stick.
Instrument parameters can be saved to USB stick.
Saved parameter can be restored from USB stick back to the instru-
ment.
Requirements to USB stick:
-
USB 2.0 supported
-
FAT 32 format (or FAT 12 or FAT 16)
-
Maximum permissible memory 32 GB
Working with USB stick:
Plug the USB stick into the USB slot located next to the keypad.
>Store NivuSoft<
>All<
All readings saved in the internal memory
are transmitted to USB stick in txt-format.
The readings can be easily imported into
NivuSoft using the “Quick Import” function.
Summary of Contents for OCM F
Page 105: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 105 Approvals and Certificates...
Page 106: ...Instruction Manual OCM F page 106 OCM F rev 05 06 04 2018...
Page 107: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 107...
Page 108: ...Instruction Manual OCM F page 108 OCM F rev 05 06 04 2018...
Page 109: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 109...
Page 110: ...Instruction Manual OCM F page 110 OCM F rev 05 06 04 2018...