Product specification
OCM F - rev. 05 / 06.04.2018
page 15
Controller
Three-step controller, quick close control, adjustable slide
valve position in error case
Data memory
4 MB, 64512 points, for programming and readings memory;
read-out via front-side USB stick
Storage cycle
1 minute to 1 hour
Ex approval (option)
II (2) G [Ex ib Gb] IIB
Sensor circuits
Ignition protection type Ex ib IIB
2-wire sensors per channel
Terminal no. 46…49
U
O
26.1 V
I
O
87.9 mA
P
O
574 mW (linear characteristics)
C
O
400 nF
L
O
5 mH
Flow sensors
Terminal no. 50…54
U
O
9.9 V
I
O
629 mA
P
O
6.2 W (rectangular characteristics)
C
O
5
µ
F
L
O
0.15 mH
Data circuits
RS485 galvanically connected to sensor circuit
Us
5 V
The maximum values also apply to concentrated capacitance / inductors that can be
switched on.
Table 2
Specifications
The type examination certificate can be found at the end of this manual.
Sensors
Observe the specifications of the according sensors as described in the respective instruction
manuals or technical descriptions.
Summary of Contents for OCM F
Page 105: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 105 Approvals and Certificates...
Page 106: ...Instruction Manual OCM F page 106 OCM F rev 05 06 04 2018...
Page 107: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 107...
Page 108: ...Instruction Manual OCM F page 108 OCM F rev 05 06 04 2018...
Page 109: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 109...
Page 110: ...Instruction Manual OCM F page 110 OCM F rev 05 06 04 2018...