Parameter Setting
OCM F - rev. 05 / 06.04.2018
page 83
25.5 CAL submenu “Simulation”
>Simulation<
Simulation of measurement.
After entering the PIN code select between level, velocity and medi-
um temperature by using the arrow keys >up< and >down<. Pressing
the arrow keys >left< or >right< will increase or decrease the simulat-
ed flow velocity, level or temperature value in steps of 1 cm or 0.1 °C.
Using ENTER enables to enter the desired simulation value directly.
The flow value which has been calculated by means of the simulated
readings will be indicated on the bottom line of the screen. Relays
which might have been set will switch and programmed mA outputs
supply according current values simultaneously.
In position “h” use the >up< and in position “t” use the >down< arrow
key to toggle between the relay status and the analog output signals
on the bottom line of the screen.
1
Simulated height/level
2
Simulated flow velocity
3
Simulated medium temperature
4
Calculated simulated flow value
5
Programmed relay activated by simulation
6
Analog output signals
Fig. 25-14 Simulation mode