Instruction Manual
OCM F
page 40
OCM F - rev. 05 / 06.04.2018
Parameter Setting
20
Basics of Parameter Setting
The transmitter in the background operates with the settings which have been entered at the
beginning of the parameter setting. The system will not ask to accept the modifications before
the settings or modifications have been finished.
Choosing >Save values< will prompt you to enter the system-PIN (may be abbreviated as
“PIN” within the display context).
2718
Enter the system PIN if prompted (password) here.
Share password / system-PIN only with authorised persons
Never share the system-PIN with unauthorised persons. Do not even leave the code next
to the equipment nor write it down on it.
The system-PIN protects from unauthorised access.
Single input of system-PIN within 24 hours
The system-PIN must be entered only once within 24 hours. The transmitter will save more
password-protected settings within this period without the need to re-enter the PIN.
After correctly entering the PIN the modified parameters are accepted and the unit will restart.
The OCM F is ready for operation again after approx. 20…30 seconds.
Options at the end of setting parameters:
•
Save modified parameters by pressing >save new values< and inserting the System-
PIN
•
Jumping back to the previous level using the >back< function to modify settings which
might have been forgotten (without the need to buffer previously modified settings)
•
Quit setting parameters and reject any modification by pressing >cancel<
Fig. 20-1
Screen on end of parameter setting
The transmitter will indicate if the
system-PIN
should have been
entered incorrectly
and
then waits for correct entry.
If you should have forgotten to enter the system-PIN use the >ESC< key to go back within
the menu.
Modifications concerning language, units and contrast do not require the code to be entered
as these settings influence just the way of representation and not measurement or output.
If parameter settings are not going to be modified but just verified by selecting each parame-
ter, there will also be no request at the end of the dialog.
Summary of Contents for OCM F
Page 105: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 105 Approvals and Certificates...
Page 106: ...Instruction Manual OCM F page 106 OCM F rev 05 06 04 2018...
Page 107: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 107...
Page 108: ...Instruction Manual OCM F page 108 OCM F rev 05 06 04 2018...
Page 109: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 109...
Page 110: ...Instruction Manual OCM F page 110 OCM F rev 05 06 04 2018...