Instruction Manual
OCM F
page 76
OCM F - rev. 05 / 06.04.2018
24.10 I/O submenu “Test control unit – manual mode”
WARNING
No safety locking measures active
The manual controller operation directly will access any following facility areas without any
safety locking measures. This might result in personal injuries.
Manual operation is for test purposes exclusively.
Safety precautions must be taken.
This menu is visible only if the controller has been enabled. The slide valve can be manually
opened and closed for testing purposes.
Use these keys to manually drive the slide valve.
1
Currently measured flow rate
2
Time how long activated relay remains energised manually [s]
3
Status of both relays
4
Status of three digital inputs
Fig. 24-12 Control menu for test control unit
25
Calibration and Calculation Menu (CAL)
This menu permits to adjust the level measurement, to adapt flow velocity and analog outputs
to the following system as well as to simulate relay switching actions and analog outputs.
Fig. 25-1
CAL Menu selection
25.1 CAL submenu “Level”
This menu allows to adjust level measurements.
Values from -1000…+1000 mm (39.4 in) can be entered. This adjustment is required only if
the level measurement utilises a pressure measurement cell.
Fig. 25-2
Submenu level
Summary of Contents for OCM F
Page 105: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 105 Approvals and Certificates...
Page 106: ...Instruction Manual OCM F page 106 OCM F rev 05 06 04 2018...
Page 107: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 107...
Page 108: ...Instruction Manual OCM F page 108 OCM F rev 05 06 04 2018...
Page 109: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 109...
Page 110: ...Instruction Manual OCM F page 110 OCM F rev 05 06 04 2018...