Setup Iridium Logger Through CONNECT
Satellite/iridium transmission is more expensive than other traditional forms of communication (e.g., cellular,
radio, Wifi). Therefore, it is imperative that the end-user only transmit the most useful diagnostic and sensor
parameters. The CONNECT software can set these parameters and their log/transmission interval before the
data logger transfers data to the web. Thus, outlining the proper sensor configuration through the CONNECT
software will reduce overall data transfer.
Note: Disconnect the iridium antenna before performing this process to ensure no communication with WQData
LIVE.
Establish Connection with Logger & Complete Sensor Detection
1. Download the CONNECT software on a PC or laptop
2. Remove one blank sensor plug from an 8-pin port (i.e., P0, P1, or P2) for each sensor connected to the X2.
Connect all sensors to the desired ports.
Note: Only one RS-232 sensor can be plugged into each P0, P1, or P2 port. Ensure that all SDI-12
and RS-485 sensors have unique addresses.
3. Connect to the X2 data logger using a
cable.
Note: Ensure the data logger receives 12VDC power from either the internal batteries on a NexSens
buoy system or the AC power adapter from a NexSens UW6-USB-485P-DC cable.
4. For new systems (i.e., no prior sensor detection), the data logger will automatically begin a sensor detection
based on its default enabled scripts. Review the article below to ensure the desired sensor scripts are
available and enabled:
Note: A new sensor detection will be required if the appropriate sensor scripts were disabled
during the automatic detection. New automatic sensor detections occur after erasing the log data
and the current sensor configuration.
a.
b.
5. After ~10 minutes (depending on the number of scripts enabled), the X2 data logger will have finished its
sensor detection.
Check the sensor configuration on the X2 using the DATA tab to review the detected sensors and
parameters.
Upload the latest data in the DAQ tab to see if the device has taken a sensor reading.
Read the last Data Acquisition
Note: Ensure to gather and evaluate a few sensor readings before continuing. It is imperative that accurate
diagnostic and sensor readings are shown before pushing the setup to the web.
44
Summary of Contents for X2-CB
Page 11: ...Select sensors from drop down list 11...
Page 17: ...Cellular X2 C to WQData LIVE 17...
Page 18: ...Iridium X2 I to WQData LIVE 18...
Page 19: ...Radio to Cellular X2 RC to WQData LIVE 19...
Page 48: ...Iridium X2 data flow 48...
Page 94: ...Rotate the antenna bushing counterclockwise to remove the antenna 94...
Page 108: ...13 Remove the Main SD Card Main SD Card Location 108...