107
SC?
Description
RS-485 controller address map query.
Syntax
SC
?
Remarks
This command is used to query the list of all controllers on an RS-485 network.
Response
Value
Description
Value
Integer
32-bit value. Default = 0
Bit#
Value
Description
0
0
The scan process did not find any address conflicts
0
1
The scan process found at least one address conflicts
1
0
There is no controller with address 1 on the network
1
1
There is a controller with address 1 on the network
31
0
There is no controller with address 31 on the network
31
1
There is a controller with address 31 on the network
Bits 1—31 are one-to-one mapped to controller addresses 1—31. The bit value is set to 1 only
when there are no conflicts with that address.
For example,
if the master controller determines that there are unique controllers at addresses
1,2, and 7 and more than one controller at address 23, this query will return 135. The binary
representation of 135 is 10000111. Bit #0 = 1 implies that the scan found at least one address
conflict during last scan. Bit #1,2, 7 = 1 implies that the scan found controllers with addresses
1,2, and 7 that do not conflict with any other controller.
Example
SC?
(get controller address map)
135
(reply; see above description)
See Also
SA, SC, SD?
Summary of Contents for New Focus Picomotor 8743-CL
Page 1: ...Picomotor Controller Driver User s Manual Model 8743 CL...
Page 10: ...x Preface...
Page 12: ...xii Preface Device Address Window 70 Internet Browser View 71 Command Syntax Diagram 73...
Page 59: ...59 Cycle Tab Cycle Motion...
Page 60: ...60 Terminal Tab Command Line Terminal...
Page 64: ...64 Options Window Connection Options Setup...
Page 65: ...65 Properties Window USB and Ethernet Properties...
Page 66: ...66 Controller Scan Scan RS 485 LAN...
Page 67: ...67 Address Conflict Resolution Window Address Conflict Resolution...
Page 70: ...70 Device Address Window...