![Newport New Focus Picomotor 8743-CL User Manual Download Page 127](http://html1.mh-extra.com/html/newport/new-focus-picomotor-8743-cl/new-focus-picomotor-8743-cl_user-manual_1679963127.webp)
127
7
Appendix
7.1
Non Axis-Specific Error Messages
0, NO ERROR DETECTED
There are no error messages in the error buffer
3, OVER TEMPERATURE SHUTDOWN
The temperature inside the controller has exceeded 85
o
C. Under this condition, the controller
automatically stops any motion that is in progress and changes the status LED color to solid RED.
When the temperature goes below 75
o
C, the controller changes the status LED color to solid
GREEN. Motion related commands can once again be issued in this state.
6, COMMAND DOES NOT EXIST
The command sent to the controller is invalid. Please check the command syntax.
7, PARAMETER OUT OF RANGE
The parameter sent along with a command is out of range.
9, AXIS NUMBER OUT OF RANGE
The axis number specified is out of range.
10, EEPROM WRITE FAILED
Unable to save the controller’s settings in its non-volatile memory.
11, EEPROM READ FAILED
Unable to restore the controller’s settings from its non-volatile memory.
37, AXIS NUMBER MISSING
The command sent to the controller requires an axis number prefix. Please check the command
syntax.
38, COMMAND PARAMETER MISSING
The command sent to the controller requires a valid parameter. Please check the command syntax.
Summary of Contents for New Focus Picomotor 8743-CL
Page 1: ...Picomotor Controller Driver User s Manual Model 8743 CL...
Page 10: ...x Preface...
Page 12: ...xii Preface Device Address Window 70 Internet Browser View 71 Command Syntax Diagram 73...
Page 59: ...59 Cycle Tab Cycle Motion...
Page 60: ...60 Terminal Tab Command Line Terminal...
Page 64: ...64 Options Window Connection Options Setup...
Page 65: ...65 Properties Window USB and Ethernet Properties...
Page 66: ...66 Controller Scan Scan RS 485 LAN...
Page 67: ...67 Address Conflict Resolution Window Address Conflict Resolution...
Page 70: ...70 Device Address Window...