Hygiene and cleaning
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
81
5 WARNING!
Patient cross-contamination and infection of the technician:
Always wear disposable gloves. Between patients, always exchange
disposable gloves, clean
EasyOne Air
, and disinfect hands. / Electric
shock, failure of
EasyOne Air
, and false diagnosis: Make sure that no fluid
penetrates the
EasyOne Flow Tube
holder or the inside of
EasyOne Air
while
cleaning. To clean
EasyOne Air
, do not fully dip
EasyOne Air
into any fluid.
To clean
EasyOne Air
and to disinfect your hands after each patient, do the
following:
a)
Use a soft cloth with a cleaning solution according to the list of cleaning
solutions under
the display of
EasyOne Air
.
b)
Put down
EasyOne Air
.
c)
Take off or change the disposable gloves and disinfect your hands
before you put on new disposable gloves.
9.2
List of cleaning solutions for
EasyOne Air
Do not use just any cleaning solution for
EasyOne Air
. Only use the listed
cleaning solutions.
CAUTION
ELECTRIC SHOCK, FAILURE OF
EASYONE AIR
, AND FALSE DIAGNOSIS
Make sure that no fluid penetrates the inside of
EasyOne Air
while cleaning.
To clean
EasyOne Air
, do not immerse
EasyOne Air
in any fluid.
To clean
EasyOne Air
from environmental dirt, you can use a damp cloth.
After each patient, clean
EasyOne Air
. You can use a soft cloth with a cleaning
solution according to the following list.
The cradle may be cleansed with the same cleaning solution.
• 70 percent isopropyl alcohol solutions or
• Solutions that contain formaldehyde (only for use outside the USA) or
• Solutions that contain chlorine (e.g. Sani-Cloth Germicidal Disposable
Wipes, EPA Reg# 9480-4)