Calibration check
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
121
7
Choose
>
> Single flow
,
Multi flow
, or
Linearity
and confirm the
message that is displayed.
8
Wait for the baseline to be set.
9
Pull the piston all the way out as slow or as fast as required by the flow level
that is displayed.
10
Push the piston all the way in again with the same speed.
11
To end the maneuver, pull the piston briefly again.
If the maneuver was not acceptable, you need to repeat steps
9
to
11
.
If you have chosen
Single flow
before and the maneuver was
acceptable, the calibration check is completed.
If you have chosen
Multi flow
or
Linearity
before and the maneuver was
acceptable, proceed with step
12
.
12
If you have chosen
Multi flow
or
Linearity
before and the previous
maneuver was acceptable, perform the next maneuvers according to the
messages that are displayed.
If a maneuver was not acceptable, you need to repeat steps
9
to
11
.
If all maneuvers were acceptable, the calibration check is completed.