Servicing and troubleshooting
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
124
List of quality messages and quality grades, 73
When
EasyOne Air
is switched on,
you see the following message on
the display: “Self-test failed”
The spirometer is possibly defective.
Turn the spirometer off and on. Try again.
If you receive the same message again,
contact your
EasyOne
dealer or the
ndd
Servicing Department.
When
EasyOne Air
is switched on,
the message “device self test
error #20” appears on the
touchscreen.
An
EasyOne Flow Tube
was inserted while
EasyOne Air
was switched on.
Wait for
EasyOne Air
to finish starting up
before you insert an
EasyOne Flow Tube
.
The
EasyOne Flow Tube
was not inserted
correctly.
Insert the
EasyOne Flow Tube
as
displayed in
Every time you switch
EasyOne Air
on, you are prompted to enter
Date etc.
The internal battery of
EasyOne Air
is
defective.
Consult your
EasyOne
dealer or the
ndd
Servicing Department.
When you start a test, you see the
following message: “Please insert
EasyOne Flow Tube
correctly”
The
EasyOne Flow Tube
is not correctly
positioned.
Ensure the
EasyOne Flow Tube
is inserted
correctly and fully pushed in.
EasyOne Air
is outside of ±3.5%
when conducting the calibration
check.
The
EasyOne Flow Tube
is not correctly
positioned.
Insert the
EasyOne Flow Tube
correctly.
You have not used an
ndd
adapter.
Use the
EasyOne Flow Tube
cal check
adapter.
There are leaks in the syringe connection. Check the connections.
The specified syringe volume does not
correspond to the actual syringe volume.
Choose the correct syringe volume under
“Configuration”.
The curve is missing on the
printout.
The ink cartridge of your printer is empty. Replace the cartridge.
In the configuration of your
EasyOne Air
,
a black and white printer is selected, but
you actually use a color printer.
Go to “Configuration”, then “Report
Settings” and select the right printer.
When printing a report, the printer
prints meaningless characters, or
the printer does not respond.
A wrong printer type has been selected in
the settings.
Set the correct printer.
The printer cable is not correctly
connected or is defective.
Switch off the spirometer and printer.
Check all plug connections.
The printer is not switched on or is not
ready.
Ensure that the printer is switched on and
also has paper. Switch the printer off and
back on again.
EasyOne Air
is not correctly positioned on
the base unit.
Insert
EasyOne Air
correctly into the base
unit.
If
EasyOne Flow Tube
is not
inserted correctly or if the
EasyOne Flow Tube
detection is
not working correctly.
An
EasyOne Flow Tube
was inserted while
EasyOne Air
was switched on the
instrument, or the
EasyOne Flow Tube
was
not inserted correctly.
Try again with the
EasyOne Flow Tube
inserted correctly. If you receive the same
message again, contact your
EasyOne
dealer or the
ndd
Servicing Department.
Wireless connection to external
device is not working correctly.
Bluetooth is not enabled,
or distance is too far,
or Bluetooth is already connected to
another device.
Enable Bluetooth (configuration),
make sure device is close enough (within
5 meters or 16.4 feet),
check setup.
Problem
Possible cause
Solution
Troubleshooting solutions