Performing spirometry
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
74
To end the test
1
To remove the
EasyOne Flow Tube
, do the following:
a)
Grip
EasyOne Air
with both hands, with the backside of
EasyOne Air
facing towards you and with the
EasyOne Flow Tube
pointing up.
b)
Press against the protruding lip of the
EasyOne Flow Tube
with both
thumbs and push the
EasyOne Flow Tube
out.
c)
Do not touch the
EasyOne Flow Tube
anywhere else than at the
protruding lip.
d)
Drop the
EasyOne Flow Tube
directly into a waste container.
2 WARNING!
Patient cross-contamination and infection of the technician:
Always wear disposable gloves. Between patients, always exchange
disposable gloves, clean
EasyOne Air
, and disinfect hands. / Electric
shock, failure of
EasyOne Air
, and false diagnosis: Make sure that no fluid
penetrates the
EasyOne Flow Tube
holder or the inside of
EasyOne Air
while
cleaning. To clean
EasyOne Air
, do not fully dip
EasyOne Air
into any fluid.
To clean
EasyOne Air
and to disinfect your hands after each patient, do the
following:
a)
Use a soft cloth with a cleaning solution according to the list of cleaning
solutions under
the display of
EasyOne Air
.
b)
Put down
EasyOne Air
.
c)
Take off or change the disposable gloves and disinfect your hands
before you put on new disposable gloves.
3
If you want to print the report immediately, choose
.
A progress bar is displayed that indicates the progress of the print-out.