A.1
Electromagnetic Compatibility (EMC)
A.1.1
General
A.1.1.1
Environment
See:
Intended use of the EasyOne Air spirometer, 2
A.1.1.2
EMC conformance
EasyOne Air
is EMC-tested in conformity with the requirements of
IEC 60601-1-2:2007 3
rd
edition (see the following tables) and
IEC 60601-1-2:2014 4
th
edition (according clause 7 and 8.9, tables 4 to 9).
EasyOne Air
is a medical device that requires special safety precautions and
must be installed and placed in operation in accordance with the information
given in tables below.
A.1.2
Safety information
See:
Safety information regarding electromagnetic compatibility, 14
A.1.3
Compliant cables and accessories
WARNING
INCREASED EMISSIONS OR DECREASED IMMUNITY
The use of accessories, transducers and cables other than those specified
may result in increased emission or decreased immunity performance of the
equipment or system.
The table below lists cables, transducers, and other applicable accessories
with which the manufacturer claims EMC compliance.
Accessories that do not affect EMC compliance are not included in the
following list.
USB cable micro to type A, shielded, 1.8 m
USB cable micro to type B, shielded, 1.8 m
ndd
power supply 5.0 V DC
Accessories with which the manufacturer claims EMC compliance