Safety information
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
16
Excessive heat
Prolonged exposure to direct sunlight, for example in cars, or exposure to oth-
er sources of heat may cause the battery to explode.
Do not store in places that can exceed temperatures of 50°C or 122°F.
Remove the battery and store the battery in the original packaging.
Network connection
Connection of
EasyOne Connect
to an IT network that includes other equip-
ment could result in previously unidentified risks to patients, operators, or
third parties.
Subsequent changes to the IT network could introduce new risks and require
additional analysis.
Changes to the IT network include:
• changes in the IT network configuration,
• connection of additional items to the IT network,
• disconnecting items from the IT network,
• update of equipment connected to the IT network, and
• upgrade of equipment connected to the IT network.
It is your responsibility to identify, analyze, evaluate, and control these risks.
IEC 80001-1:2010 provides guidance to address these risks.
2.3
Safety information regarding electromagnetic compatibility
Influence by HF
surgical equipment
Electrical devices with a high RF power output during intended use (e.g. High
frequency (HF) surgical equipment) should not be operated in parallel with
EasyOne Air
.
Portable wireless
communications
equipment
Portable wireless communications equipment such as wireless home network
devices, mobile phones, cordless telephones and their base stations, walkie-
talkies, etc. can affect
EasyOne Air
.
Keep a distance of at least 30 cm (12 inches) to any part of
EasyOne Air
.
Increased emissions
or decreased
immunity
Accessories, transducers, and cables other than those specified by the manu-
facturer or replacement parts for internal components may result in increased
emissions or decreased immunity of
EasyOne Air
.
Proximity to other
equipment
EasyOne Air
should not be used adjacent to or stacked with other equipment.
Should close proximity or stacking be unavoidable, however, the configura-
tion in which it is used needs to be watched closely to ensure that the equip-
ment is functioning normally.
Interruption of the
USB connection
Electrical fast transient coupling to the USB cable or electrostatic discharge on
the device may cause an interruption of the USB connection, which is indicat-
ed to the user. After reconnecting, the USB connection will work as expected.
There is no harm because the data transmission is secured.
See Appendix