Performing spirometry
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
74
2
Fully insert the
EasyOne Flow Tube
into the
EasyOne Flow Tube
holder as
depicted below, but keep the partly unwrapped wrapper on the mouthpiece
of the
EasyOne Flow Tube
.
The shape of the
EasyOne Flow Tube
and the
EasyOne Flow Tube
holder
guide the orientation of the
EasyOne Flow Tube
. You can only insert the
EasyOne Flow Tube
fully if it is positioned correctly.
When you insert the
EasyOne Flow Tube
, avoid direct sunlight as it can
interfere with the automatic recognition of the
EasyOne Flow Tube
.
3
On the main menu of
EasyOne Connect
, choose
Patients
.
4
Select the patient that you want to test.
5
At the bottom of the screen, choose
Test
.
6
Choose the test that you want to perform.
7
If the environment conditions screen is displayed, enter the current values
and choose
Confirm >>
.
For the FVL test, the MVV test, and the SVC test, you must enter the
environment temperature within 1°C or 1.8°F accuracy.
If the patient has already performed a test on the same day, a selection
window is displayed. Proceed with step
8
.
If the patient has not performed a test on the same day, the screen of
the selected test is displayed with a pop-up window that asks you to block
the
EasyOne Flow Tube
. Proceed with step
9
.