3 Functions
3.4 Setting modes
D-52/53/54/55
83
During automatic data storage, the MEAS, MODE, SET,
ENTER, and CAL DATA keys cannot be used. Data
recording time will differ ± 1 seconds from the time set by
the storage interval. If the number of stored data items
exceeds 300, data storage will stop and the error
message “ERR No. 10” will be displayed.
3.4.6 pH calibration frequency setting
If you set a number of days in the calibration frequency
setting, ERR 8 (calibration frequency error) will be dis-
played after the specified days have passed since the
last calibration.
1.
Press the SET key in the Measurement mode to
select the Calibration Interval Setting mode.
2.
Pressing the ENTER key toggles the calibration
frequency setting between ON and OFF.
3.
Pressing the MODE key toggles between CH1
and CH2 (D-53 ).
4.
Specify a numerical value using the
▲
and
▼
keys.
Setting range: 1 to 400 days