2 Taking Measurements
2.6 Measuring ORP
36
HORIBA
●
Measuring ORP
1.
Immerse the electrode all the way in the sample
solution.
For accurate measurements, be sure to immerse the
electrode in the sample at least three centimeters.
2.
Press the MODE key, once.
The ORP Instantaneous Value Measurement
screen will appear.
3.
Press the MEAS key with the initial screeen
displayed.
"HOLD" will blink on the display until the reading
stabilizes.
When the value stabilizes, "HOLD" will stop blinking and
will be displayed continually. The indicated value will
remain displayed, and ORP measurement is completed.
Refer to the “ Criteria for judging stability” page 23 for the
criteria for judging the stability of the readout.