4 RS-232C communications
4.3 On-line operation commands
D-52/53/54/55
109
●
Start ion calibration command format (D-53 )
・
This command is valid when on-line, in ion
measurement, or while “HOLD” is displayed.
・
If the start calibration command is sent while in the
Measurement mode, the previous calibration values
will be cleared.
Setting ranges for calibration values
When the units are mg/L,
µ
mol/L, mmol/L or
µ
mol/L:
・
0.00 - 9.99
([SP][SP]X or [SP]X.X are also possible.)
・
10.0 - 99.9
([SP]XX is also possible.)
[SP]100 - [SP]999
Prefix for the unit of the calibration value
to be set
0: -; 1: m; 2:
μ
The set unit determines which unit is
used, either mol/L or g/L.
Header
Command code (Ion calibration start)
Calibration value being set