6 Maintenance and Troubleshooting
6.5 Troubleshooting
D-52/53/54/55
161
●
ERR No.05 Electrode sensitivity error (DO)
Explanation
If there was something wrong with the DO calibration,
re-calibrate after taking the appropriate measures listed
below.
Cause
How to solve problem
The settings
(temperature,
correction of
salinity
concentration,
or air-pressure
correction) are
wrong.
Reconfirm each setting
(temperature, correction of salinity
concentration, and air-pressure
correction).
There is liquid
on the DO tip
membrane.
(when
conducting air
calibration)
Let the electrode sit until the liquid
evaporates or remove the liquid
using soft tissue paper, making sure
not to scratch the membrane.
There is
something
wrong with the
standard
solution.
(when
conducting
standard
solution
calibration)
Prepare new zero and span
standard solutions.
The stirring is
inappropriate.
Stir the solution appropriately (at a
constant speed, between 1000 and
1500 rpm). (Make sure the stirrer
does not emit heat.)
The electrode is
defective.
If the DO tip is dirty, clean it.
If the DO tip membrane is damaged
or the DO tip is worn out, replace it.