4 RS-232C communications
4.3 On-line operation commands
D-52/53/54/55
111
●
Start DO calibration command format (D-55 )
・
This command is valid when on-line, when DO
measurement or calibration are on HOLD.
・
After calibration is finished,
END will be displayed
and the calibration coefficient will be stored in the
memory.
●
Clear calibration command format
・
This command is valid when on-line, when “HOLD” is
being displayed for ION or pH calibration.
・
When a pH electrode is being calibrated, the pH
standard calibration value is cleared.
・
When an ION electrode is being calibrated, the ION
standard calibration is cleared.
・
The calibration value of the channel being displayed
currently will be cleared.
Header
Calibration method to be set (one digit)
0: Air calibration
1: Standard solution calibration
Command code (DO calibration start)
Header
Command code (Calibration clear)
Channel number (1 or 2)