6 Maintenance and Troubleshooting
6.5 Troubleshooting
170
HORIBA
Check this point
As shown in the diagram, use a jumper wire or bent
paper clip to short the meter by touching both the center
pin and some metal part in the electrode connector.
If the flaching measured value disappears when this
done, the meter is nomal.
●
CLR is flashing (during ION measurement)
Concentration cannot be measured correctly.
Cause
How to solve problem
The pH meter is
in default (initial-
ized) status.
Calibrate the pH meter.