6 Maintenance and Troubleshooting
6.5 Troubleshooting
D-52/53/54/55
157
●
ERR No. 01 Memory error
Explanation
Data cannot be read from or written to the internal
memory.
●
ERR No. 02 Battery voltage low
Explanation
The battery has insufficient voltage.
The measured value cannot be guaranteed when ERR
No. 02 is displayed.
Cause
How to solve problem
The pH meter
does not start
operating cor-
rectly even after
the power is
turned ON.
Take the battery from the pH meter,
and disconnect the AC adapter.
Then press the ON/OFF key for
about 10 seconds.
The internal IC
is defective.
Seek repairs at your nearest retail
outlet or HORIBA service station.
Cause
How to solve problem
The battery
voltage is low.
(Battery voltage:
2.2V or less)
Replace the dry-cell battery.