6 Maintenance and Troubleshooting
6.2 ION electrode maintenance
D-52/53/54/55
143
6.2
ION electrode maintenance
Refer to the electrode operation manual for how to take
care of each kind of electrode.
●
ION electrodes
* Change in the electric potential of the electrode (25ºC)
when the ION concentration changes by a factor of
10.
The above electrodes are subject to change without
notice.
Maintenance of the ION electrodes listed in the above
table is described on the next page.
Ion to be
measured
Ion
type
Slope
( * )
Measurement
range
Elect-
rode
model
Compatible
tip model
Reference
solution
Potassium
K
+
+1
+58 mV 0.04 – 39,000
mg/L
6582
-10C
7682
3.33 mol/L
NaCl
Calcium
Ca
2+
+2
+29 mV 0.4 – 40,080
mg/L
6583
-10C
7683
3.33 mol/L
KCl (#300)
Chloride
CI
-
-1
-59 mV 0.4 – 35,000
mg/L
6560
-10C
7660
1 mol/L
KNO
3
Fluoride
F
-
-1
-59 mV 0.02 – 19,000
mg/L
6561
-10C
7661
3.33 mol/L
KCl (#300)
Nitrate
NO
3
-
-1
-55 mV 0.06 – 62,000
mg/L
6581
-10C
7681
1 mol/L
KCl
Ammonia
NH
3
+1
-59 mV 0.1 – 1,000
mg/L
5002
-10C
Included
internal
solution
NH
4
Cl
Summary of Contents for D-25
Page 2: ......
Page 11: ...CONTENTS VIII HORIBA ...
Page 31: ...1 Overview of the Meter 1 9 Using the protective cap D 53 54 55 20 HORIBA ...
Page 75: ...2 Taking Measurements 2 9 Dissolved oxygen DO measurement D 55 64 HORIBA ...
Page 139: ...4 RS 232C communications 4 5 Communication example using the HyperTerminal 128 HORIBA ...