Connecting Acid Dosing Unit APE
See also Chapter 3.9 Acid injection, page 17.
CAUTION
Acid from the condensate
Acid burns may result from weakly acidic liquids from the
condensate.
►
If you come into contact with acid, wash the area im-
mediately using a lot of water.
►
Note the safety data sheet for phosphoric acid (10%).
►
Consider to connect a container to the condensate
out port
NOTE
Phosphoric acid (10%) is not included in the scope of
delivery.
Phosphoric acid (10 %) must be procured by the
operator.
►
Check the remaining contents of the acid tank
before measurement.
►
Connect a collection container to the condensate outlet.
►
Plug the AUX-connector into the analyser.
The AUX-connection cable ensures that no phosphoric acid (10 %) is
injected into the analyser when the pumps are switched off.
►
Plug the couplings of the connecting hose into the connection
nozzle of the Acid Dosing Unit APE (2x outlet to device).
►
Plug the couplings of the connecting hose into the connection
nozzle of the analyser.