After the set time from start monitoring after zeroing has elapsed,
monitoring starts.
Monitoring status shows: Monitoring is running
The analyser continuously displays the newest H
2
0 value [ppm].
As soon as the permitted H
2
0 deviation [ppm] is exceeded, the
status changes to: Overrun was detected.
If the allowed duration of deviation [min] is exceeded, the status
changes to: A second overrun during the allowed duration of
deviation was detected. Perform new zeroing!
A message appears.
►
You have the possibility to perform a new zeroing.
►
You have the possibility to ignore the message and reset the water
monitoring. The water monitoring starts again.
►
You have the possibility to ignore the message and stop the water
monitoring.