USER MANAUL
MGA
prime Q
MRU GmbH, D-74172 Neckarsulm
22 /
117
►
Touch
The display turns.
4.7.
Set the analyser to standby mode
The analyser is protected in standby mode. You cannot take
measurements in standby mode.
The actuators, for example pumps and gas coolers, are switched off as
far as possible. Note that NDIR is not switched off. After you have left
the standby mode, the analyser will start up again.
The analyser will display a message that a zeroing may be useful.
►
Touch the Context menu (19) on the display.
A selection list appears.
►
Touch Standby.
A message appears.
►
Press Yes to activate standby mode.
The sensors are purged.
The analyser is set to standby mode.
►
Press MEASUREMNT MODE? to restart the analyser.
A message appears.
►
Press Yes to exit the standby mode.
A message appears.