USER MANAUL
MGA
prime Q
MRU GmbH, D-74172 Neckarsulm
88 /
117
12.2.
Calibrating NDIR reference points
The following steps show the calibration procedure for an NDIR cuvette
in the analyser:
The calibration menu can be opened in the menu NDIR 1096.
►
Open the menu.
►
To enter the calibration menu of the cuvette, a PIN must be entered.
See also chapter 11 Entering Pin code, Page 86.
The password is available from the producer on request.
►
Go to the menu NDIR 1096.
►
Wait until zeroing is successfully completed.
►
Press the filed ADJUSTMENT.
►
In this menu all gases are listed to calibrate the respective reference
point.
NDIR 1096
ADJUSTMENT
3
1
2