USER MANAUL
MGA
prime Q
MRU GmbH, D-74172 Neckarsulm
113 /
117
►
Use remote control via Android device
NOTE
Please note that due to the increased computing
power, the CPU temperature will increase. If the tem-
perature should become too high, the VNC server
switches off automatically.
18.6.
CxHy Calculation
The calculation for CxHy is:
•
if CH
4
and C
3
H
8
are present: CxHy = CH
4
+ C
3
H
8
.
•
if only one HC - channel is available, CxHy is equal to it.
18.7.
General Instructions for the heated hose line
For the save operation of the sample line it is important that it is in-
stalled correctly. This includes the following points:
•
The bending radius as given below is maintained,
•
The entire heated hose line is rolled out before put into operation,
•
Both ends of the heated hose line are connected correctly,
•
For thermal and mechanical reasons, it shall be supported by a
cable duct along the line,
•
The heated hose line shall be supported close to the analyser cab-
inet and probe housing, for example by using a separate stilt,
•
There are no extreme temperature variations along the line, as the
line’s temperature is regulated depending on a sensor placed in
the mid of the line.