When reading stable, you can press ‘ENTER’ key, meter automatically calculates new TDS
factor, calibration finishes. Press ‘CANCEL’, meter stops the calibration of TDS factor.
4.9.5. Salinity Measurement
Before measure salinity, you must refer to the chapter Selection of Conductive Electrode to
confirm the conductive electrode to be used in measurement, setup suitable cell constant. Under
starting state, if you have already selected conductive measurement mode parameter, such as
conductivity parameter, you can start measurement directly. Otherwise, you should enter into
‘Setup Measurement Mode’
, select conductive measurement unit to measure. (
Please refer to
‘Set Measurement Mode’ for details
).
For the display, measurement procedure and operating, please refer to chapter Conductivity
Measurement.
Chart 1 KCL standard solution for measurement cell constant
Electrode constant (l/cm)
0.01
0.1
1
10
KCL solution approximate concentration
(mol/L)
0.001
0.01
0.01or 0.1
0.1or 1
Chart 2 Compound of standard solution
Approximate
concentration
(mol/L)
Volume concentration KCL(g/L) solution(20
℃
in air)
1
74.2650
0.1
7.4365
0.01
0.7440
0.001
dilute100mL 0.01mol/Lsolution to 1L
Chart 3 Relationship between KCL solution approximate concentration and conductivity
Appro.
Temp.
conc. (mol/L)
15.0
℃
18.0
℃
20.0
℃
25.0
℃
30.0
℃
1
12120
97800
101700
111310
131100
0.1
10455
11163
11644
12852
15353
0.01
1141.4
1220.0
1273.7
1408.3
1687.6
0.001
118.5
126.7
132.2
146.6
176.5