Press ‘CUSTOM’ and select ‘Create New Mode’, ‘Delete Ion Mode’ or ‘Amend Ion Mode’ to
custom ion mode.
You can only start concentration measurement when you select correct ion mode. If you select
different ion mode, it will cause incorrect measurement result.
For example
, if you want to measure concentration of sodium ion, then you should select ‘Na
+
’
ion mode in ‘Set Ion Mode’ and start to measure concentration of sodium ion.
4.5.3. Set System Time(Date & Time)
Press ‘START’, select ‘Date & Time’ and press ‘ENTER’ key. Meter enters into setup ‘Set System
Time’ mode. You can set present system time.
On the upper side of the window is present time and
date
At the bottom is the corresponding time button
including ‘year’, ‘month’, ‘day’, ‘hour’, ‘minute’,
‘second’. It is displayed as follows:
If you want to amend time, you can press ‘EDIT’, press ‘↑ ↓’, ‘←→’ and move cursor to the item
which is needed to be amend, then press ‘EDIT’ and input time figure.
For example,
if you want to set present month, you can operate as follows:
Press ‘↑ ↓’, ‘←→’ and move the cursor to ‘Month’ item,
Press ‘EDIT’ again, meter pop up input window, you can input the month. When input
finishes, press ‘ENTER’ key to quit the input window.
Your can setup other time according to the same procedure.
When all time items are amended, you can press ‘ENTER’ key to finish and press ‘CANCEL’ key
to quit ‘Set System Time’ mode and return to starting state.
4.5.4. Set Manual Temperature
When temperature sensor socket is connected with temperature sensor, meter will automatically
set the temperature value. Otherwise, current temperature value is set by user.
Press ‘START’(F1), select ‘Manual Temp’ and press
‘ENTER’ key, meter enters into ‘Manual Temperature’ mode.
Press ‘EDIT’(F3) to amend temperature value. You can input
temperature value according to actual need manually.