93
Flush out the various water systems everyday before treatment.
Turn on the main switch and follow the procedure described below.
(1)Using the Flushing Device
*
Set up the flushing bowl and flush out the various systems.
Refer to section [5] 7. (1) 5) in this manual.
(2)Not Using the Flushing Device
Be sure to start from step 1) and follow the procedure
exactly; otherwise the flushing procedure will not be
performed efficiently.
The times for flushing each system (see chart to the right)
are predicated on the amount of water in the system.
Longer flushing times are recommended to include the
water in the main line leading to the treatment unit.
1)Auto Filler Flushing
a)Put a cup in place and fill it.
b)Empty the cup and refill it. Repeat 8 times.
c)The manual cup switch can also be used. In this case hold the switch
down for 1 minute.
2)Three-way Syringe Flushing
Do both the dentist’s and the assistant’s at the same time.
a)Hold the nozzles over the basin and run water through the syringes
for at least 1 minute.
5. Flushing Out the Water System
CAUTION
Flush out the water system in the way described below at least once a day
before treatment. This is especially important if the unit has not been used
for a day or more such as after a holiday.
>>The quality of water that has been sitting in the warmer tank or in the tubes will
degrade, in quality and must be flushed out.
[6]Maintenance 5.
*: Option
ADVICE
CHECK
Estimated Flushing Times
System Time
Auto Filler
Three-way Syringe 1min.
Air Turbine
Micromotor 2min.
Ultrasonic Scaler 3min.